- Laborwassersysteme
- |
- Geno™ II



Geno™ II
Geno™ II Deionised (DI) water purification systems for laboratories and hospitals. Produces Type 2 General Laboratory Grade Water.
- Type 2 DI water
- RO sanitisation process
- Simple operation with easy-change consumables
- Visual alarms to alert users when the system outside of set parameters, e.g., water quality
- Mount on a bench, a wall, or on top of a reservoir
- Intuitive, icon-based, interactive user interface with touchscreen
- Integrated leak detector and boost pump
- Choice of storage reservoir size
Technische Daten

Geno™ II
Make-Up Flow Rate (@15°C/59°F)
≥25L/hr
Rejection Rate
≥98%
Rate from Tank
3 - 7 L/min
Typical TOC Reduction
96%
Recommended Daily Usage
160 L/day
Pressure
0.1 - 6 Bar/1.45 - 86 PSI
Conductivity
<1400μS/cm
Temperature
4 - 30°C/39.2 - 86°F
Hardness as CaCO3
<400 ppm
Free Chlorine
<2 ppm
Collid Index
<3 SDI
pH Value
4 - 10
Iron (Total Fe)
<0.1 ppm
Dimensions (H x W x D)
500 x 490 x 290 mm/19.7" x 19.3" x 11.5"
Installed Weight
23 kg/50.7 lbs
Power Supply
100 - 240 v ~; 50/60 Hz; 3.0A (max)
Power Supply Outbound
24 VDC, 5.0 A, 120 W (max)
Operating Temperature
4°C - 35°C max)/39.2°F - 95°F (max)
Drain Type
Sink/Tundish/Standard Up-Stand with Suitable Trap
Drain Flow